What is the sensor field-of-view (FOV) of Caregiver?

The sensor in Caregiver has a field of view of 16° full cone angle as shown in the figure. (The sensor element is recessed approximately 1" from the from the end of the thermometer.) This applies to BODY and SURFACE modes. At 1" from the end of the thermometer, it will measure the [...]

By |2025-04-07T18:40:08+00:00April 7th, 2025||Comments Off on What is the sensor field-of-view (FOV) of Caregiver?

I See One of the Following Error Codes on the LCD “LO HI E1 E2 E4 E5”. What does it mean?

Error or Warning Codes may be displayed on the LCD.  Here is the interpretation: Warning and Error Codes E1- The operational (ambient) temperature is BELOW specification range E2 – The operational (ambient) temperature is ABOVE specification range E4 – Sensor / circuit error (usually, this means the device needs to be repaired) Lo – [...]

By |2025-04-07T18:35:55+00:00April 7th, 2025||Comments Off on I See One of the Following Error Codes on the LCD “LO HI E1 E2 E4 E5”. What does it mean?